Table 2 Comparison of detection performance between AGSCS and G-test for 198 clinical serum samples (Enzyme micro-plate reader)

From: A novel aptamer-G-quadruplex/hemin self-assembling color system: rapid visual diagnosis of invasive fungal infections

Definite diagnosis

AGSCS

Total

G-test

Total

Positive

Negative

Positive

Negative

Positive

76

6

82

71

11

82

Negative

12

104

116

20

96

116

Total

88

110

198

91

107

198

  1. The 198 serum sample from human were detected using AGSCS and G-test. The sensitivity of AGSCS and G-test were 92.68% and 86.58%, the specificity of both were 89.65% and 82.75%, respectively. The accuracy of AGSCS and G-test were 90.9% and 84.34% respectively. The AGSCS showed better detectability for patients with IFI than G-test (2 = 3.936, p = 0.0473).