Table 9 Sensitivity, specificity and accuracy of the 4 Widal brand for anti-H antibodies of 91 random serum samples of the 150 clinically diagnosed typhoid fever cases using IgM anti-LPS ELISA as a reference test at three cut-off values

From: The dilemma of widal test - which brand to use? a study of four different widal brands: a cross sectional comparative study

Widal Brand

sensitivity

specificity

accuracy

Using a cut-off 1/80

Remel

94.02%

8.33%

71.42%

Biosystems

91.04%

16.66%

71.42%

Biotech

86.56%

20.83%

69.23%

Dialab

73.13%

45.83%

65.93%

Using a cut-off 1/160

Remel

83.58%

20.83%

67.03%

Biosystems

80.59%

29.16%

67.03%

Biotech

73.13%

54.16%

68.13%

Dialab

55.22%

91.66%

64.83%

Using a cut-off 1/320

Remel

56.71%

79.16%

62.63%

Biosystems

47.76%

74.07%

57.14%

Biotech

35.82%

91.66%

50.54%

Dialab

19.40%

100.00%

40.65%