Skip to main content

Table 4 Distribution of the results of 91 serum samples using the 4 Widal brands for anti-O antibodies in relation to their corresponding IgM anti-LPS ELISA results using a cut-off 1/320

From: The dilemma of widal test - which brand to use? a study of four different widal brands: a cross sectional comparative study

Widal brand

anti-O titer

≥ 1/320

ELISA (91)

Total

  

Positive (67)

Negative (24)

 
  

Number

%

Number

%

 

Remel (91)

Negative

27

40.29

24

100.00

51

 

Positive

40

59.7

0

0.00

40

Biosystems (91)

Negative

36

53.73

24

100.00

60

 

Positive

31

46.26

0

0.00

31

Biotech (91)

Negative

43

64.17

24

100.00

67

 

Positive

24

35.82

0

0.00

24

Dialab (91)

Negative

65

97.01

24

100.00

89

 

Positive

2

2.98

0

0.00

2